Main
Ecap Concept
Project Requirements
Shear Force Analysis
Finite Element Analysis
Final Design
Op Manual
Contact
References
 
Operations Manual

Operations Manual
Equal Channel Angular Pressing




Introduction:

Hello and thank you for your recent purchase of an Equal Channel Angular Press (ECAP). The ECAP is a fabricated die with a 45-degree angle change that makes a sample of material under go shear deformation. This shear deformation takes this sample of material and causes severe plastic deformation in the sample. Unlike other systems the ECAP that you have purchased is designed to deform samples of copper while submerged in liquid nitrogen. This is done because of the Hall-Petch Phenomenon, which says the colder the temperature at which the sample is deformed the smaller the grain sizes will be. Making the material increase in strength.



Figure 1: View of all the parts required:

In figure 1, above you will see all the parts you should have in order to assemble the ECAP. To setup the ECAP please refer to the assembly section. Figure 2 on the next page will show you what the solid view of the parts look like.


Figure 2: Solid view of all parts required

In order to operate the die we will be using a 500-ton press shown in figure 3, with at least a 20-inch height workspace. Please refer to assembly to see how to setup the die and press.



Figure 3: 500-ton press

When the sample of material (copper) is done with a pass through the die it will need to be resized. This is going to be done with a separate machine called a resizing die. The resizing die will return the sample back to its original size so that it can be ready for another pass. The sample must be marked so you can remember the orientation of the sample after each pass. The sample will need to be rotated 90 degrees after every pass. So be sure to mark the sample so the orientation of the sample can be remembered. The samples will under go at 8 passes through the die. This depends on the type of material being used. This was determined from room temperature dies that found the materials did not benefit from more than 8 passes through the die.

Remember these instructions are for a sample of copper.

Equal Channel Angular Processing

Safety Equipment/Tools Needed:

Eyes:

• Chemical safety goggles and face shield are recommended when handling N2 liquid.

Skin:  

• Wear loose fitting thermal insulated leather protective gloves.
• Wear long sleeve shirt for handling liquid.
• Wear safety shoes if handling cylinders.

Respiratory:


• Use a self-contained breathing apparatus or positive pressure airline with mask in oxygen-deficient atmospheres.
• Respirators will not function.

Special Precautions:

Ventilation:
1. Use local exhaust and
2. Mechanical ventilation.

Storage:

1. Store and use with adequate ventilation.
2. Containers should be stored upright and firmly secured to prevent falling or being knocked over.
3. Protect containers from physical damage; do not drag, roll, slide, or drop.
4. Use a suitable hand truck for container movement.
5. Liquid containers will vent nitrogen if internal pressure builds up, so these containers should be stored in well-ventilated areas.

Other Tools Needed:

• Standard Crowbar or Large Flat-headed screwdriver.
• Protection shield to place in front of 500-ton press during experimentation.
• Tongs (minimum of 18 inches) capable of retrieving processed sample from reservoir.
• Wrench (11/4in.)
• Re-sizing Die.


Total Assembly/Component List




1:) Top Bolster Plate
2:) Piston Cylinder
3:) Right Die Half
4:) Left Die Half
5:) 3/4in x 9in. Bolt (11 total)
6:) 11/4 in. Nut (11 total)
7:) Insulated Bath
8:) Extrusion Channel
9:) Catch pan
10:) 2-inch Bottom Bolster plate (not shown)


Assembly of the apparatus:
In order to insure that the die and press work in unison together, below is an ordered list of assembly instructions for the operator of the die. It also gives proper instructions of the placement in the press.

Note: the part in the numbered diagram will be denoted by a “(#)”

Die Halves Connection:

1. Find a flat level table to begin assembly (check with level).
2. Remove the left die half (4) from the shipping box/storage
3. Remove the right die half (3) from the shipping box/storage
4. Place the right die half (3) and the left die half (4) parallel to each other as shown in figure 1.
5. After aligning the two dies, press them together to make a solid block as seen in figure #. Make sure the holes and channel are aligned on both halves to insure proper sealing.
6. Insert the eleven 9” bolts (5) into the boltholes of the left die and extending through the right die. Completely insert all bolts (5) before nuts are screwed on.
7. Once all bolts are completely inserted, screw the nuts (6) on one by one, holding the head of the bolt still with a wrench (Do not tighten bolts and screws until all are snug against the die).
8. Slowly tighten each bolt ¼ turn each until the two die become snug together. When finished it should look like figure 4.



Figure 4: Assembled die halves

Die Placement in Bath & Press Placement
1. After die is fully assembled, place the bath on the bottom bolster plate, with the catch pan facing front and hanging off the edge and then, place the die in the bath (7). Be sure to align the bottom of the die with the 1” alignment plates coming up from the bath floor see in figure #.
2. Once the die is in place in the bath, and on top of the bottom bolster plate, the bath and bolster plate are ready to be set up on the press.
3. Make sure the press is separated in the vertical direction to its maximum (20”).
4. Attach the bottom bolster plate (10), with the bath and die on top, to the bottom of the press (secure bolts down)
5. Attach and secure the top bolster plate (1) to the top of the press. Four bolts will be used for this process.
6. Pour the liquid nitrogen into the bath and allow the dies to reach the equilibrium temperature.
7. Slide the square end of the piston cylinder (2), in the slot in the top bolster plate.
8. Be sure that the bath is stable and aligned and that the piston (2) is aligned with the extrusion channel (8).


Sample Size

The samples that the dies are able to accommodate are .75 inch outside diameter, and 6 inches long. The die was designed for use with copper samples, the use of a sample of a different material may cause damage to the die and personal injury.


Setup
The following sections illustrate how to properly setup the die assemblies and press to process a sample.

500 Ton Press Orientation

The first step to setup up the assembly begins with properly setting up the press. First, the press must be visually inspected for any visible defects. Next, making sure the press is properly lubed; set it to the fully open position.

Bolster Plates

The 2-inch aluminum, bottom bolster plate (10) must be set on a table or cart for the bath and die to be placed on top. The 1-inch top bolster plate (1) must be bolted to the top of the press opening with the notch for the piston on the top.

Bath Placement

The liquid nitrogen bath container (7) can next be placed on the bottom bolster plate and oriented so that the catch pan (9) faces the front and is hanging over the edge of the bottom bolster plate.

Die Insertion

Take the left (4) and right (3) side die halves, and place them together. Insert the bolts (5) through both dies and place the nuts (6) on the end. The bolts should be tightened finger tight, and then another ¼ turn afterwards (refer to section “Die Halves Connection” for more detailed instruction). Then, place the dies into the bottom bolster plate and transfer over to the press. This assembly will be heavy, so a rolling cart should be used to move it closer to the press (refer to section “Die Placement in Bath & Press Placement“ for more detailed instruction).

Adding Liquid Nitrogen

Wearing safety goggles and gloves for protection, carefully poor the liquid nitrogen into the bath till it is almost full up to the top. Next, place the lid on top and be sure to wait a sufficient amount of time to allow the die to reach an equilibrium temperature. The bath will need to be refilled periodically as the liquid nitrogen boils off.

Sample Placement

Before the samples are to be pressed, they must be cooled in a liquid nitrogen tank. Remove the lid and place a sample from the liquid nitrogen tank into the hole in the top of the die. Then, put the lid back in its place.


Piston Arrangement

Now, the piston (2) can be inserted into the top bolster plate. With the square base on the topside, slide it into the notch in the bolster plate and line it up with the hole in the lid and die.


Experimental Procedure

1. The first thing to do is to follow the setup instructions to assemble the ECAP and the press. Remember to mark the sample to be able to remember its orientation
2. Place the sample of copper, with a diameter of 0.75inches and length of 5 inches, into the die channel. The lid might be off when placing the sample into the die channel See figure 5 on next page. Caution use gloves our tongs to place sample in channel and then replace lid if not on. Allow sample to come to equilibrium temperature; should take about 1hr.



Figure 5: Showing sample going into the die

3. If piston cylinder is not aligned then please align them now.
4. Once the piston cylinder is aligned with channel you are ready to begin pressing the sample. To ensure that the sample is not pressed to fast the minimum press time should be 2 minutes.
NOTE: When performing the ECAP there should be two samples pressed one after the other. The first sample will be pressed out but the second will remain in the die. It will be help full to have one extra sample to be used to push the last sample out with out using one of the samples that have already been through the die 8 passes.
5. Once the sample has been pressed through remove lid and use tongs to reach in to the catch pan and grab sample and remove it out of the liquid nitrogen.
6. The samples should still be marked and able to tell how it was oriented before, so after resizing it can be reoriented by 90 degrees.
7. The sample will go into the resizing die to have the diameter of the sample corrected back to 0.75 inches.
8. After resizing the sample will be placed back into the die channel. But remember the sample must be rotated by 90 degrees from its previous position (until it has gone 360 degrees). The sample must pass through the die in sets of four so that the sample is deformed evenly getting all 360 degrees. So by sending the sample through 8 passes the whole sample will have made two complete revolutions.
9. Steps 5-8 should be repeated to the desired number of passes (8)
10. There will be one sample left in the die that you care about. So place the third extra sample in die channel and press out the second sample that you care about leaving the extra sample that does not matter in the die.
11. Follow disassembly to take apart die and remove last sample.

Disassembly of the Die
This is a brief explanation of the complete disassembly, also noting the trouble-shooting for jamming of the copper sample in the extrusion channel.

1. Allow ample time for the liquid nitrogen to evaporate and for the assembly to reach room temperature.
2. Remove piston (2) from the top bolster plate (1) notch.
3. Remove the Styrofoam lid from the insulated bath.
4. Carefully remove the bath and die assembly along with the bottom bolster plate (10) from the press.
5. Remove the die assembly from the insulated bath and place on a level table.
6. Loosen and remove the 9” bolts (5) and nuts (6) from the die assembly.
7. Insert screwdriver/crowbar in the V-notch between the two die halves and pry them apart.
8. After this point the copper sample left in the die is then removed from the extrusion channel.

Note: This process will be used for the case of a sample getting jammed in the extrusion channel also before another pass.